On a neutron detector based on TlInSe2 crystals intercalated with a lithium isotope

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

The possibility of increasing the sensitivity of a neutron detector based on a TlInSe2 crystal by introducing the 6Li isotope into this crystal is investigated. Introduction of the isotope is conducted by the method of electrochemical intercalation from aqueous and nonaqueous solutions of LiCl and also from the melt of the LiCl–KCl eutectic. It is shown that intercalation by the electrochemical method from a LiCl solution in propylene carbonate, performed along the “c” axis of the crystal (along strong bond chains) is efficient. The attained concentration of introduced lithium amounts to (1–1.2) × 1021 cm–3, which increased the sensitivity of the detector by approximately four times. Good agreement between the calculated and experimental values of the detector sensitivity is demonstrated.