On a neutron detector based on TlInSe2 crystals intercalated with a lithium isotope


Цитировать

Полный текст

Открытый доступ Открытый доступ
Доступ закрыт Доступ предоставлен
Доступ закрыт Только для подписчиков

Аннотация

The possibility of increasing the sensitivity of a neutron detector based on a TlInSe2 crystal by introducing the 6Li isotope into this crystal is investigated. Introduction of the isotope is conducted by the method of electrochemical intercalation from aqueous and nonaqueous solutions of LiCl and also from the melt of the LiCl–KCl eutectic. It is shown that intercalation by the electrochemical method from a LiCl solution in propylene carbonate, performed along the “c” axis of the crystal (along strong bond chains) is efficient. The attained concentration of introduced lithium amounts to (1–1.2) × 1021 cm–3, which increased the sensitivity of the detector by approximately four times. Good agreement between the calculated and experimental values of the detector sensitivity is demonstrated.

Об авторах

I. Alekseev

Joint Institute for Nuclear Research

Автор, ответственный за переписку.
Email: aliv2@yandex.ru
Россия, Dubna, 141980

E. Goremychkin

Joint Institute for Nuclear Research

Email: aliv2@yandex.ru
Россия, Dubna, 141980

N. Gundorin

Joint Institute for Nuclear Research

Email: aliv2@yandex.ru
Россия, Dubna, 141980

A. Petrenko

Joint Institute for Nuclear Research

Email: aliv2@yandex.ru
Россия, Dubna, 141980

I. Sashin

Joint Institute for Nuclear Research

Email: aliv2@yandex.ru
Россия, Dubna, 141980

Дополнительные файлы

Доп. файлы
Действие
1. JATS XML

© Pleiades Publishing, Ltd., 2017

Согласие на обработку персональных данных

 

Используя сайт https://journals.rcsi.science, я (далее – «Пользователь» или «Субъект персональных данных») даю согласие на обработку персональных данных на этом сайте (текст Согласия) и на обработку персональных данных с помощью сервиса «Яндекс.Метрика» (текст Согласия).