Effect of Sample-Shape Imperfection on Uncertainty in Measurements of the Thermal-Conductivity by the Laser-Flash Method

Abstract

The effect of the geometric shape of the samples on the uncertainty in measurements of the thermal conductivity coefficient of materials by the laser-flash method is investigated. A mathematical model simulating the process of measuring the thermal conductivity of samples made of graphite, Mg2Si0.4Sn0.6, and bismuth telluride by the laser-flash method is created in the Comsol Multiphysics software. Cylindrical samples with plane-parallel sides, samples shaped as a truncated cylinder, as well as parallelepiped-shaped samples with a square base are investigated. It is shown that the uncertainty in the measurements does not exceed 2% for the samples with plane-parallel sides and sizes up to 12.7 mm. For the samples shaped as a truncated cylinder with a diameter of 3 mm and at an inclination angle of φ = 1.5°, the measurement uncertainty does not exceed 3%. With increasing sample diameter and angle φ, the measurement uncertainty increases significantly.

About the authors

A. V. Asach

ITMO University

Email: vladdisslav.fomin@yandex.ru
Russian Federation, St. Petersburg, 197101

G. N. Isachenko

ITMO University; Ioffe Institute

Email: vladdisslav.fomin@yandex.ru
Russian Federation, St. Petersburg, 197101; St. Petersburg, 194021

A. V. Novotelnova

ITMO University

Email: vladdisslav.fomin@yandex.ru
Russian Federation, St. Petersburg, 197101

V. E. Fomin

ITMO University

Author for correspondence.
Email: vladdisslav.fomin@yandex.ru
Russian Federation, St. Petersburg, 197101

K. L. Samusevich

ITMO University

Email: vladdisslav.fomin@yandex.ru
Russian Federation, St. Petersburg, 197101

I. L. Tkhorgevskii

ITMO University

Email: vladdisslav.fomin@yandex.ru
Russian Federation, St. Petersburg, 197101


Copyright (c) 2019 Pleiades Publishing, Ltd.

This website uses cookies

You consent to our cookies if you continue to use our website.

About Cookies