A comparison of electrospray tandem mass spectra of some sialic acid derivatives: Ion trap and high resolution QqToF mass spectrometers


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

Using O-acetyl-N-acyl derivatives of O-methyl sialoside methyl esters, it was shown that an ion trap and a hybrid analyzer (linear quadrupole–time-of-flight analyzer, reflectron) give comparable, though not identical secondary mass spectra for the [M + Na]+ and [M + K]+ ions. A parallel use of an ion trap and a hybrid QqToF instrument gives information about the fragmentation pathways of ions of sialic acid derivatives under collisional activation. In this case, the sequence of fragmentation may be established using an ion trap, whereas a QqToF instrument offers a possibility of revealing the elemental composition of fragment ions quickly and unequivocally.

About the authors

A. O. Chizhov

Zelinsky Institute of Organic Chemistry

Author for correspondence.
Email: chizhov@ioc.ac.ru
Russian Federation, Moscow, 119991

A. M. Shpirt

Zelinsky Institute of Organic Chemistry

Email: chizhov@ioc.ac.ru
Russian Federation, Moscow, 119991

N. Y. Savel’eva

Zelinsky Institute of Organic Chemistry; Higher Chemical College

Email: chizhov@ioc.ac.ru
Russian Federation, Moscow, 119991; Moscow, 125047

L. O. Kononov

Zelinsky Institute of Organic Chemistry

Email: chizhov@ioc.ac.ru
Russian Federation, Moscow, 119991


Copyright (c) 2016 Pleiades Publishing, Ltd.

This website uses cookies