Using of Fractional Separation for the Analysis of MoO3 by Atomic Emission Spectrometry with Inductively Coupled Plasma and Electrothermal Vaporization
- 作者: Medvedev N.S.1, Malitskii A.V.1,2, Tsygankova A.R.1,2, Saprykin A.I.1,2
-
隶属关系:
- Nikolaev Institute of Inorganic Chemistry
- Novosibirsk State University
- 期: 卷 73, 编号 6 (2018)
- 页面: 533-538
- 栏目: Articles
- URL: https://journals.rcsi.science/1061-9348/article/view/183051
- DOI: https://doi.org/10.1134/S1061934818060072
- ID: 183051
如何引用文章
详细
The introduction of various forms of molybdenum into an inductively coupled plasma was studied in the vaporization of solutions from a graphite tube. A temperature program is selected that enables the separated vaporization of analytes and molybdenum (matrix) for atomic emission spectrometry with inductively coupled plasma and electrothermal vapoization (ETV–ICP–AES) analysis. The limits of detection for analytes in the ETV–ICP–AES analysis of molybdenum trioxide are evaluated using the fractional separation of analytes and the matrix.
作者简介
N. Medvedev
Nikolaev Institute of Inorganic Chemistry
Email: saprykin@niic.nsc.ru
俄罗斯联邦, Novosibirsk, 630090
A. Malitskii
Nikolaev Institute of Inorganic Chemistry; Novosibirsk State University
Email: saprykin@niic.nsc.ru
俄罗斯联邦, Novosibirsk, 630090; Novosibirsk, 630090
A. Tsygankova
Nikolaev Institute of Inorganic Chemistry; Novosibirsk State University
Email: saprykin@niic.nsc.ru
俄罗斯联邦, Novosibirsk, 630090; Novosibirsk, 630090
A. Saprykin
Nikolaev Institute of Inorganic Chemistry; Novosibirsk State University
编辑信件的主要联系方式.
Email: saprykin@niic.nsc.ru
俄罗斯联邦, Novosibirsk, 630090; Novosibirsk, 630090
补充文件