Transmission for a Quadrupole Mass Filter with Dipolar Excitation


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

Mass peaks are calculated using the numerical simulation of ion trajectories in the linear quadrupole with the dipole excitation(s). Using this data, diagrams for transmission versus resolution are calculated both for the ideal quadrupole field and for the quadrupole field with round rods. It is shown that, using the dipole excitation mode, resolution may be increased from R0.1 = 1000 to R0.1 = 2200 with 20% QMF transmission in an ideal case, and, for the round rods, the dipole excitation along the isoline βy = 0.01 enables an increase in resolution R0.1 from 500 to 800 at a 20% level of transmission.

作者简介

K. Seregin

Ryazan State University

Email: n.konenkov@rsu.edu.ru
俄罗斯联邦, Ryazan, 390000

N. Konenkov

Ryazan State University

编辑信件的主要联系方式.
Email: n.konenkov@rsu.edu.ru
俄罗斯联邦, Ryazan, 390000

A. Berdnikov

Institute for Analytical Instrumentation, Russian Academy of Sciences

Email: n.konenkov@rsu.edu.ru
俄罗斯联邦, St. Petersburg, 190103