Determination of 2,4-dichlorophenoxyacetic acid in cereals by fluorescence polarization immunoassay


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

A fluorescence polarization immunoassay (FPIA) procedure is developed for the determination of 2,4-dichlorophenoxyacetic acid in cereal grains using a Sentry-200 portable device. We synthesized tracers, that is, antigens labeled with fluorescein derivatives, based on two fluorescent compounds and two chlorophenoxyacids. The tracer synthesized from 2,4-dichlorophenoxyacetic acid and 4-aminomethylfluorescein is found to be optimal for the FPIA of 2,4-dichlorophenoxyacetic acid. The limit of detection for 2,4-dichlorophenoxyacetic acid in flour was 40 ng/g; the analytical range is from 80 to 1000 ng/g. The percentage of recovery was 85 ± 7% at the relative standard deviation (RSD) 2–10%.

Sobre autores

A. Boroduleva

Department of Chemistry

Autor responsável pela correspondência
Email: boroduleva.anna@gmail.com
Rússia, Moscow, 119991

S. Eremin

Department of Chemistry

Email: boroduleva.anna@gmail.com
Rússia, Moscow, 119991


Declaração de direitos autorais © Pleiades Publishing, Ltd., 2016