A Dissolved Oxygen Analyzer with an Optical Sensor


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

A serial analyzer of dissolved oxygen, Expert-009, based on the optical detection method is created. The sensitive element of the sensor is a polymer film with a dye—a metal complex of fluorine-substituted porphyrin—distributed in it. The method is based on the quenching of the phosphorescence of the dye by molecular oxygen, which the analyzer recounts to the concentration value by the calibration curve method. The metrological characteristics of the analyzer are determined.

About the authors

N. K. Zaitsev

Moscow Technological University (MITHT)

Email: melnikovsoft@mail.ru
Russian Federation, Moscow, 119571

V. I. Dvorkin

Moscow Technological University (MITHT)

Email: melnikovsoft@mail.ru
Russian Federation, Moscow, 119571

P. V. Melnikov

Moscow Technological University (MITHT)

Author for correspondence.
Email: melnikovsoft@mail.ru
Russian Federation, Moscow, 119571

A. E. Kozhukhova

Moscow Technological University (MITHT)

Email: melnikovsoft@mail.ru
Russian Federation, Moscow, 119571


Copyright (c) 2018 Pleiades Publishing, Ltd.