Peculiarities of the Mass Spectrometric Detection of Anthocyanins in High-Performance Liquid Chromatography


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

The dependence of the degree of fragmentation of anthocyanin “molecular” ions on the cone voltage of a mass spectrometric electrospray ionization detector was studied. It was found that the voltage required for the fragmentation of 50% of original “molecular” ions, Ef(0.5), increased with the number of glycoside residues. The fragmentation of glycosides proceeds with the removal of the entire residue regardless of their structures. In the case of 3,5-diglycosides, two types of fragment ions formed due to the loss of glycosidic residues from different positions; the ratio of their peak intensities is reciprocal to the ratio of the masses of residues eliminated. The values of Ef(0.5) for monoglycosides (190 V), diglycosides (229 V), triglycosides (267 V), and for some acylated cyanidin-3,5-diglycosides are determined. These results were given for the gradient separation of anthocyanins by reversed-phase HPLC in aqueous acetonitrile containing 10 vol % of formic acid using quadrupole mass spectrometric detection.

About the authors

V. I. Deineka

Institute of Engineering, Technologies and Natural Sciences

Author for correspondence.
Email: deineka@bsu.edu.ru
Russian Federation, Belgorod, 308015

A. N. Sidorov

Institute of Engineering, Technologies and Natural Sciences

Email: deineka@bsu.edu.ru
Russian Federation, Belgorod, 308015

A. N. Chulkov

Belgorod Branch

Email: deineka@bsu.edu.ru
Russian Federation, Belgorod, 308027

L. A. Deineka

Institute of Engineering, Technologies and Natural Sciences

Email: deineka@bsu.edu.ru
Russian Federation, Belgorod, 308015


Copyright (c) 2017 Pleiades Publishing, Ltd.

This website uses cookies

You consent to our cookies if you continue to use our website.

About Cookies