Design of a Single-Crystal Diffractometer for the PIK Reactor


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

The scheme of arrangement of a four-circle diffractometer on a thermal neutron beam from the HEC-9 channel of the PIK reactor is proposed. Calculations are performed using analytical and numerical approaches. The setup parameters are optimized: the resolution, the size of the focusing monochromator and its curvature in the vertical and horizontal directions, the efficiency of filters, and the monochromatic neutron-flux density on the sample, with which the maximum intensity is achieved in the detector. It is shown that the beam intensity at the position of the detector is a more important parameter than the intensity at the sample position during device optimization. The device characteristic obtained will allow studying the atomic and magnetic structure of a wide class of crystals.

作者简介

A. Bykov

Konstantinov Petersburg Nuclear Physics Institute, National Research Centre “Kurchatov Institute”

编辑信件的主要联系方式.
Email: redi87@bk.ru
俄罗斯联邦, Gatchina, Leningrad oblast, 188300

O. Smirnov

Konstantinov Petersburg Nuclear Physics Institute, National Research Centre “Kurchatov Institute”

Email: redi87@bk.ru
俄罗斯联邦, Gatchina, Leningrad oblast, 188300

补充文件

附件文件
动作
1. JATS XML

版权所有 © Pleiades Publishing, Ltd., 2019