Design of a Single-Crystal Diffractometer for the PIK Reactor
- Авторы: Bykov A.A.1, Smirnov O.P.1
-
Учреждения:
- Konstantinov Petersburg Nuclear Physics Institute, National Research Centre “Kurchatov Institute”
- Выпуск: Том 13, № 5 (2019)
- Страницы: 898-907
- Раздел: Article
- URL: https://journals.rcsi.science/1027-4510/article/view/196464
- DOI: https://doi.org/10.1134/S1027451019050252
- ID: 196464
Цитировать
Аннотация
The scheme of arrangement of a four-circle diffractometer on a thermal neutron beam from the HEC-9 channel of the PIK reactor is proposed. Calculations are performed using analytical and numerical approaches. The setup parameters are optimized: the resolution, the size of the focusing monochromator and its curvature in the vertical and horizontal directions, the efficiency of filters, and the monochromatic neutron-flux density on the sample, with which the maximum intensity is achieved in the detector. It is shown that the beam intensity at the position of the detector is a more important parameter than the intensity at the sample position during device optimization. The device characteristic obtained will allow studying the atomic and magnetic structure of a wide class of crystals.
Ключевые слова
Об авторах
A. Bykov
Konstantinov Petersburg Nuclear Physics Institute, National Research Centre “Kurchatov Institute”
Автор, ответственный за переписку.
Email: redi87@bk.ru
Россия, Gatchina, Leningrad oblast, 188300
O. Smirnov
Konstantinov Petersburg Nuclear Physics Institute, National Research Centre “Kurchatov Institute”
Email: redi87@bk.ru
Россия, Gatchina, Leningrad oblast, 188300
Дополнительные файлы