X-ray Absorption Spectroscopy Analysis of Heavy Metals by Means of a Silicon Detector
- 作者: Osadchii S.M.1, Petukhov A.A.1, Dunin V.B.2
-
隶属关系:
- All-Russian Research Institute of Physical-Technical and Radiotechnical Measurements
- Joint Institute for Nuclear Research
- 期: 卷 13, 编号 4 (2019)
- 页面: 683-689
- 栏目: Article
- URL: https://journals.rcsi.science/1027-4510/article/view/196396
- DOI: https://doi.org/10.1134/S1027451019040116
- ID: 196396
如何引用文章
详细
A Si-PIN detector with longitudinal arrangement of the silicon wafer for detecting X-ray quanta with an energy of up to 100 keV is designed. Its resolution is estimated. An X-ray absorption spectroscopy method for the detection of heavy metals at the K-absorption edges is implemented using the designed detector. The results of measuring the thicknesses of lead and gold foils placed behind a shield by the X-ray absorption spectroscopy method and with a micrometer gauge are compared. The detection limits of heavy metals behind steel or aluminum shields are determined. The measurement error of the metal-foil thicknesses is estimated.
作者简介
S. Osadchii
All-Russian Research Institute of Physical-Technical and Radiotechnical Measurements
编辑信件的主要联系方式.
Email: osm@vniiftri.ru
俄罗斯联邦, Mendeleevo, Moscow oblast, 141570
A. Petukhov
All-Russian Research Institute of Physical-Technical and Radiotechnical Measurements
Email: osm@vniiftri.ru
俄罗斯联邦, Mendeleevo, Moscow oblast, 141570
V. Dunin
Joint Institute for Nuclear Research
Email: osm@vniiftri.ru
俄罗斯联邦, Dubna, Moscow oblast, 141980
补充文件