Integrating Devices in the Measurement Circuits of Induction Magnetometers


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

New structural designs of integrating devices by means of which the precision of integration of electrical signals in the range of ultra-low frequencies 2·10–3–3 Hz may be increased are presented. Practical designs of second-order integrators and their transfer functions are considered. Results of calculations that confirm an increase in the precision of integration of signals by a second-order integrator by comparison with a first-order integrator as a function of the frequency of the device are presented.

About the authors

M. N. Kuligin

Murom Institute, Branch of Vladimir State University

Author for correspondence.
Email: itpu@mivlgu.ru
Russian Federation, Murom


Copyright (c) 2019 Springer Science+Business Media, LLC, part of Springer Nature

This website uses cookies

You consent to our cookies if you continue to use our website.

About Cookies