Highly Sensitive Chromatography Mass Spectrometry Determination of Impurities in High-Purity Monogermane Using Adsorption Capillary Column with Carbon Sorbent
- 作者: Krylov V.A.1,2, Chernova O.Y.1, Sozin A.Y.1
-
隶属关系:
- Devyatykh Institute of Chemistry of High-Purity Substances
- Lobachevsky State University of Nizhny Novgorod (NNSU)
- 期: 卷 53, 编号 14 (2017)
- 页面: 1386-1390
- 栏目: Analysis of Substances
- URL: https://journals.rcsi.science/0020-1685/article/view/158353
- DOI: https://doi.org/10.1134/S0020168517140102
- ID: 158353
如何引用文章
详细
It is shown that the employment of a quartz capillary column with carbon adsorbent provides an effective separation of impurities of permanent gases, carbon dioxide, nitrous oxide, C1–C2 hydrocarbons, and carbon dioxide in monogermane. The effect of the injected volume of monogermane specimen on the determination of ethane and carbon dioxide is studied. Chromatography mass spectrometry detection limits of impurities are (1–10) × 10–6 mol %. The detection limit of the limiting impurity represented by ethane is decreased by the factor of seven as compared to the published data.
作者简介
V. Krylov
Devyatykh Institute of Chemistry of High-Purity Substances; Lobachevsky State University of Nizhny Novgorod (NNSU)
编辑信件的主要联系方式.
Email: k658995@mail.ru
俄罗斯联邦, Nizhny Novgorod, 603950; Nizhny Novgorod, 603950
O. Chernova
Devyatykh Institute of Chemistry of High-Purity Substances
Email: k658995@mail.ru
俄罗斯联邦, Nizhny Novgorod, 603950
A. Sozin
Devyatykh Institute of Chemistry of High-Purity Substances
Email: k658995@mail.ru
俄罗斯联邦, Nizhny Novgorod, 603950
补充文件