Nonaqueous suspensions in total reflection X-ray fluorescence analysis of solids
- Authors: Sharanov P.Y.1, Alov N.V.1, Zolotov Y.A.1
-
Affiliations:
- Moscow State University
- Issue: Vol 467, No 1 (2016)
- Pages: 91-93
- Section: Chemistry
- URL: https://journals.rcsi.science/0012-5008/article/view/153618
- DOI: https://doi.org/10.1134/S0012500816030101
- ID: 153618
Cite item
Abstract
It has been suggested to use suspensions of solid samples in viscous liquids for the determination of the elemental composition by the total reflection X-ray fluorescence (TRXF) method. Two liquids—ethylene glycol and glycerol—have been studied as a dispersion phase. It turned out that the sedimentation stability of suspensions in these liquids is higher than in water. During 5 min of sedimentation, the suspended analyte concentration changes by 10–15 rel. % in nonaqueous media and by up to 50 rel. % in aqueous media. It has been determined that the repeatability of measurements in nonaqueous suspensions is 3–5 rel. % against 10–15 rel. % for aqueous suspensions. It has been demonstrated that ethylene glycol and glycerol can be efficiently used for preparing samples for the TRXF method.
About the authors
P. Yu. Sharanov
Moscow State University
Email: n_alov@mail.ru
Russian Federation, Moscow, 119991
N. V. Alov
Moscow State University
Author for correspondence.
Email: n_alov@mail.ru
Russian Federation, Moscow, 119991
Yu. A. Zolotov
Moscow State University
Email: n_alov@mail.ru
Russian Federation, Moscow, 119991