Nonaqueous suspensions in total reflection X-ray fluorescence analysis of solids
- 作者: Sharanov P.Y.1, Alov N.V.1, Zolotov Y.A.1
-
隶属关系:
- Moscow State University
- 期: 卷 467, 编号 1 (2016)
- 页面: 91-93
- 栏目: Chemistry
- URL: https://journals.rcsi.science/0012-5008/article/view/153618
- DOI: https://doi.org/10.1134/S0012500816030101
- ID: 153618
如何引用文章
详细
It has been suggested to use suspensions of solid samples in viscous liquids for the determination of the elemental composition by the total reflection X-ray fluorescence (TRXF) method. Two liquids—ethylene glycol and glycerol—have been studied as a dispersion phase. It turned out that the sedimentation stability of suspensions in these liquids is higher than in water. During 5 min of sedimentation, the suspended analyte concentration changes by 10–15 rel. % in nonaqueous media and by up to 50 rel. % in aqueous media. It has been determined that the repeatability of measurements in nonaqueous suspensions is 3–5 rel. % against 10–15 rel. % for aqueous suspensions. It has been demonstrated that ethylene glycol and glycerol can be efficiently used for preparing samples for the TRXF method.