Nonaqueous suspensions in total reflection X-ray fluorescence analysis of solids


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

It has been suggested to use suspensions of solid samples in viscous liquids for the determination of the elemental composition by the total reflection X-ray fluorescence (TRXF) method. Two liquids—ethylene glycol and glycerol—have been studied as a dispersion phase. It turned out that the sedimentation stability of suspensions in these liquids is higher than in water. During 5 min of sedimentation, the suspended analyte concentration changes by 10–15 rel. % in nonaqueous media and by up to 50 rel. % in aqueous media. It has been determined that the repeatability of measurements in nonaqueous suspensions is 3–5 rel. % against 10–15 rel. % for aqueous suspensions. It has been demonstrated that ethylene glycol and glycerol can be efficiently used for preparing samples for the TRXF method.

作者简介

P. Sharanov

Moscow State University

Email: n_alov@mail.ru
俄罗斯联邦, Moscow, 119991

N. Alov

Moscow State University

编辑信件的主要联系方式.
Email: n_alov@mail.ru
俄罗斯联邦, Moscow, 119991

Yu. Zolotov

Moscow State University

Email: n_alov@mail.ru
俄罗斯联邦, Moscow, 119991

补充文件

附件文件
动作
1. JATS XML

版权所有 © Pleiades Publishing, Ltd., 2016