Fluorimeter with Reflecting Optical Elements for Detecting of Simultaneous Reactions in a Microplate
- Authors: Chubakov V.P.1, Chubakov P.A.1
-
Affiliations:
- Institute of Automation and Electrometry, Siberian Branch
- Issue: Vol 54, No 5 (2018)
- Pages: 484-489
- Section: Optical Information Technologies
- URL: https://journals.rcsi.science/8756-6990/article/view/212573
- DOI: https://doi.org/10.3103/S8756699018050096
- ID: 212573
Cite item
Abstract
This paper describes the autocollimation optical scheme of a fluorimeter that allows detecting different reactions from multiple wells of a microplate at the same time. The image focusing is implemented a spherical mirror that excludes the presence of background illumination and chromatic aberration. The operation of the fluorimeter is demonstrated by measuring a signal from the microplate with 96 and 384 individual wells.