Free hemoglobin reference intervals detected with hemoglobin-cyanide method on a biochemical automatic analyzer ILAB 300 PLUS

Cover Page

Cite item

Full Text

Abstract

Objectives – to evaluate the reference intervals for free hemoglobin concentration with the help of hemoglobin-cyanide method and a biochemical automatic analyzer.

Material and methods. The reference group consisted of 120 healthy men and 120 healthy women from the Astrakhan region aged from 20 to 60 years.

The fulfilment of standard procedures during the pre-analytical pre-laboratory, pre-analytical laboratory, analytical and post-analytical stages of the study ensured the quality control for free hemoglobin evaluation.

Blood samples were taken with cubital vein puncture after applying a tourniquet (for not more than 1 minute), patient in lying position, using two-component blood sampling systems – disposable polypropylene tubes with a blood coagulation activator (Sarstedt, Germany). To obtain blood serum, blood tubes were centrifuged at 2500 rpm during 10 minutes.

The free hemoglobin in blood serum was detected by the hemoglobin-cyanide method using a ILAB 300 PLUS biochemical automatic analyzer (Laboratory Instrumentation, USA).

The data was processed using the software package Statistica 6.0 for Windows (StatSoft Inc, США).

Results. We used the classical approach and followed the strict inclusion and exclusion criteria, recommended by CLSI C 28-A3 guidelines. Gender and age differences in the content of free hemoglobin in blood serum in men and women of the Astrakhan region were not found. The reference interval of free hemoglobin in men and women of the Astrakhan region was evaluated as 0.02–0.84 g/l.

About the authors

Olga I. Murygina

Federal Center for Cardiovascular Surgery

Email: students_asma@mail.ru
ORCID iD: 0000-0002-0359-8664

physician of clinical laboratory diagnostics

Russian Federation, Astrakhan

Elena R. Zhukova

Federal Center for Cardiovascular Surgery

Email: students_asma@mail.ru
ORCID iD: 0000-0002-5314-4601

physician of clinical laboratory diagnostics

Russian Federation, Astrakhan

Olga V. Petrova

Federal Center for Cardiovascular Surgery; Astrakhan State Medical University

Author for correspondence.
Email: students_asma@mail.ru
ORCID iD: 0000-0003-3544-2266

PhD, Head of Clinical Diagnostic Laboratory

Russian Federation, Astrakhan

Dina M. Nikulina

Astrakhan State Medical University

Email: students_asma@mail.ru

PhD, Professor, Head of the Department of Biological Chemistry

Russian Federation, Astrakhan

References

  1. Kasyanova TR, Levitan BN, Kriventsev YuA, et al. The etiologic peculiarities of changes of fetal hemoglobin in patients with chronic hepatitis and liver cirrhosis. Astrahanskij medicinskij zhurnal. 2012;(4):138–140. (In Russ.) [Касьянова Т.Р., Левитан Б.Н., Кривенцев Ю.А. и др. Этиологические особенности фетального гемоглобина у больных хроническим гепатитом и циррозом печени. Астраханский медицинский журнал. 2012;7(4):138–140].
  2. Tonkoshkurova OA, Dmitriyev AI, Dmitriyeva RE. Measurement of nonerythrocytic hemoglobin of the blood serum or plasma by the hemoglobincyanide metod. Klinicheskaja laboratornaja diagnostika. 1996;2:21–22. (In Russ.) [Тонкошкурова О.А., Дмитриев А.И., Дмитриева Р.Е. Определение концентрации внеэритроцитарного гемоглобина плазмы (сыворотки) крови гемоглобинцианидным методом. Клиническая лабораторная диагностика. 1996;2:21–22].
  3. Spiridonova VA, Novikova TM, Nikulina DM, et al. Complex formation with protamine prolongs the thrombin inhibiting effect of DNA aptamer in vivo. Biochimic. 2018;145:158–162. doi: 10.1016/j.biochi.2017.09.010
  4. Petrova OV, Shabanova GR, Egorova TG. Reference intervals for the white blood cell and differential white blood cell count in the adult population at the use of the automatic hematology analyzer Sysmex XT 2000I. Gematologija i transfuziologija. 2016;61(3):153–156. (In Russ.) [Петрова О.В., Шабанова Г.Р., Егорова Т.Г. Референтные интервалы количества лейкоцитов в крови и лейкоцитарной формулы у взрослого населения при применении автоматического гематологического анализатора Sysmex XT 2000I. Гематология и трансфузиология. 2016;61(3):153–156]. doi: 10.18821/0234-5730-2016-61-3-153-1
  5. Petrova OV, Shashin SA, Tarasov DG. Reference values of platelet aggregation in impedance aggregometry with adenosine diphosphoric acid on aggregometer multiplate. Sovremennye tehnologii v medicine. 2016;8(3):100–104. (In Russ.) [Петрова О.В., Шашин С.А., Тарасов Д.Г. Референтные значения агрегации тромбоцитов при исследовании импедансным методом с аденозиндифосфорной кислотой на агрегометре Multiplate. Современные технологии в медицине. 2016;8(3):100–104. doi: 10.17691/stm2016.8.3.11
  6. Petrova OV, Urtaeva ZA, Shashin SA, et al. The reference intervals of antitrombin III in population of the Astrakanskaia oblast at application of automated coagulometers. Klinicheskaja laboratornaja diagnostika. 2016;61(7):429–432. (In Russ.) [Петрова О.В., Уртаева З.А., Шашин С.А. и др. Референсные интервалы антитромбина III у населения Астраханской области при применении автоматических коагулометров. Клиническая лабораторная диагностика. 2016;61(7):429–432]. doi: 10.18821/0869-2084-2016-61-7-429-432
  7. Chumakova SP, Shipulin VM, Urasova OI, et al. The role of uneven blood oxygen saturation parameter in development of hemolysis after on-pump coronary after bypass graft surgery. Fundamental'naja i klinicheskaja medicina. 2018;3 (1):22–29. (In Russ.) [Чумакова С.П., Шипулин В.М., Уразова О.И. и др. Роль неравномерной оксигенации крови и других условий перфузии в патогенезе гемолиза при операциях с искусственным кровообращением. Фундаментальная и клиническая медицина. 2018;3 (1):22–29].
  8. CLSI. Defining, establishing, and verifying reference intervals in the clinical laboratory; appoved guideline. CLSI Document C28-A3c Third ed. Wayne, PA: Clinical and Laboratory Standards Institute; 2008. https://clsi.org.
  9. Daly C, Liu X, Grey V, Hamid J. A systematic review of statistical methods used in constructing pediatric reference intervals. Clin Biochem. 2013;46(13-14):1220–7. doi: 10.1016/j.clinbiochem.2013. 05.058
  10. Shaw J, Cohen A, Konforte D, Binesh-Marvasti T, et al. Validity of establishing pediatric reference intervals based on hospital patient data: a comparison of the modified Hoffmann approach to CALIPER reference intervals obtained in healthy children. Clin Biochem. 2014;47(3):166–172. doi: 10.1016/j.clinbiochem.2013.11.008
  11. Sinclair L, Hall S, Badrick T. A survey of Australian haematology reference intervals. Pathology. 2014;46(6):538–43. doi: 10.1097/PAT.0000000000000148

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2019 Murygina O.I., Zhukova E.R., Petrova O.V., Nikulina D.M.

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.

Согласие на обработку персональных данных

 

Используя сайт https://journals.rcsi.science, я (далее – «Пользователь» или «Субъект персональных данных») даю согласие на обработку персональных данных на этом сайте (текст Согласия) и на обработку персональных данных с помощью сервиса «Яндекс.Метрика» (текст Согласия).