Low-Energy Electron Detection by Total-Absorption Cherenkov Spectrometer


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

The results of calibration of the total-absorption Cherenkov spectrometer based on TF-1 lead glass of a thickness 14.8X0 using a quasi-monochromatic electron beam are presented. It is found that the energy resolution is 89–10% in the electron beam energy range E = 6–285 MeV, correspondingly.

About the authors

V. I. Alekseev

Lebedev Physical Institute, Russian Academy of Sciences

Email: baskov@x4u.lebedev.ru
Russian Federation, 53 Leninskii Pr., Moscow, 119991

V. A. Baskov

Lebedev Physical Institute, Russian Academy of Sciences

Author for correspondence.
Email: baskov@x4u.lebedev.ru
Russian Federation, 53 Leninskii Pr., Moscow, 119991

V. D. Dronov

Lebedev Physical Institute, Russian Academy of Sciences

Email: baskov@x4u.lebedev.ru
Russian Federation, 53 Leninskii Pr., Moscow, 119991

A. I. L’vov

Lebedev Physical Institute, Russian Academy of Sciences

Email: baskov@x4u.lebedev.ru
Russian Federation, 53 Leninskii Pr., Moscow, 119991

A. V. Kol’tsov

Lebedev Physical Institute, Russian Academy of Sciences

Email: baskov@x4u.lebedev.ru
Russian Federation, 53 Leninskii Pr., Moscow, 119991

Yu. F. Krechetov

Joint Institute for Nuclear Research

Email: baskov@x4u.lebedev.ru
Russian Federation, Joliot-Curie St. 6, Dubna, Moscow Region, 141980

E. I. Malinovskii

Lebedev Physical Institute, Russian Academy of Sciences

Email: baskov@x4u.lebedev.ru
Russian Federation, 53 Leninskii Pr., Moscow, 119991

V. V. Polyanskii

Lebedev Physical Institute, Russian Academy of Sciences

Email: baskov@x4u.lebedev.ru
Russian Federation, 53 Leninskii Pr., Moscow, 119991

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2019 Allerton Press, Inc.