Nutation line shape for the nonstationary regime of magnetic resonance flowmeter–relaxometer


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

An upgrade of the structure of magnetic resonance flowmeter–relaxometer is proposed to improve the metrological characteristics. The improved structure of the magnetic resonance flowmeter–relaxometer is based on the modulation of static magnetic field Н0 by ac magnetic field in the region of the nutation coil. The experiments show that the measurements of liquid flow q in the proposed regime make it possible to eliminate the effect of measurement errors related to variations in amplitude and phase of the NMR signal inverted by 180° under rapid variations in quantity q and increase the measurement dynamic range of quantity q by 30%.

作者简介

V. Davydov

St. Petersburg Polytechnic University

编辑信件的主要联系方式.
Email: Davydov_vadim66@mail.ru
俄罗斯联邦, Politekhnicheskaya ul. 29, St. Petersburg, 195251

V. Dudkin

St. Petersburg Polytechnic University

Email: Davydov_vadim66@mail.ru
俄罗斯联邦, Politekhnicheskaya ul. 29, St. Petersburg, 195251

N. Myazin

St. Petersburg Polytechnic University

Email: Davydov_vadim66@mail.ru
俄罗斯联邦, Politekhnicheskaya ul. 29, St. Petersburg, 195251

补充文件

附件文件
动作
1. JATS XML

版权所有 © Pleiades Publishing, Inc., 2016