Calibration of CryojetHT and Cobra Plus Cryosystems used in X-ray diffraction studies


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

CryoJetHT (Oxford Instruments) and Cobra Plus (Oxford Cryosystems) cryosystems, which are used for sample cooling in X-ray diffraction experiments, have been calibrated. It is shown that the real temperature in the vicinity of the sample differs significantly (the deviation is as high as 8–10 K at low temperatures) from the temperature recorded by authorized sensors of these systems. The calibration results are confirmed by measurements of the unit-cell parameters of GdFe3(BO3)4 single crystal in the temperature range of its phase transition. It is shown that, to determine the real temperature of a sample, one must perform an independent calibration of cryosystems rather than rely on their ratings.

Sobre autores

A. Dudka

Shubnikov Institute of Crystallography

Autor responsável pela correspondência
Email: dudka@crys.ras.ru
Rússia, Leninskii pr. 59, Moscow, 119333

I. Verin

Shubnikov Institute of Crystallography

Email: dudka@crys.ras.ru
Rússia, Leninskii pr. 59, Moscow, 119333

E. Smirnova

Shubnikov Institute of Crystallography

Email: dudka@crys.ras.ru
Rússia, Leninskii pr. 59, Moscow, 119333

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Pleiades Publishing, Inc., 2016