Instrument function of a broadband acoustic thermometric detector


Дәйексөз келтіру

Толық мәтін

Ашық рұқсат Ашық рұқсат
Рұқсат жабық Рұқсат берілді
Рұқсат жабық Тек жазылушылар үшін

Аннотация

The instrument function of a broadband (1.6–2.5 MHz) detector that is used in acoustic thermometry has been calculated. Experimental tests have proved that measured and computed results are in agreement. The effect of the pass band characteristics and the detector’s dimension on the instrument function has been studied as well as the effect that the instrument function has on an acoustic thermometric signal that is measured by the detector. The ratio of the wavelength (for the mean reception frequency) to the detector’s radius has been shown to be the main parameter that determines the acoustic thermometric signal at distances that are typical of acoustic thermometry. For problems of localizing a heated domain, it is optimal to locate the receiver at a distance of 15–25 mm from the domain. For example, for a detector 8 mm in diameter, the width of the instrument function at a level of 0.5 of the maximum is 1.2 ± 0.1 mm in this zone.

Авторлар туралы

A. Anosov

Kotel’nikov Institute of Radio-Engineering and Electronics; Sechenov First Moscow State Medical University

Хат алмасуға жауапты Автор.
Email: anosov@hotmail.ru
Ресей, Moscow, 125009; Moscow, 119992

A. Sharakshane

Plekhanov Russian University of Economics

Email: anosov@hotmail.ru
Ресей, Moscow, 117997

A. Kazanskii

OOO AURIGA

Email: anosov@hotmail.ru
Ресей, Moscow, 117587

A. Mansfel’d

Institute of Applied Physics

Email: anosov@hotmail.ru
Ресей, Nizhny Novgorod, 603950

A. Sanin

Institute of Applied Physics

Email: anosov@hotmail.ru
Ресей, Nizhny Novgorod, 603950

A. Sharakshane

Kotel’nikov Institute of Radio-Engineering and Electronics

Email: anosov@hotmail.ru
Ресей, Moscow, 125009

Қосымша файлдар

Қосымша файлдар
Әрекет
1. JATS XML

© Pleiades Publishing, Ltd., 2016