Development of the Method of Calibration Equations for the X-Ray Fluorescence Analysis of Multicomponent Samples in the Presence of Undetectable Elements
- 作者: Garmai A.1, Oskolok K.1
-
隶属关系:
- Department of Chemistry
- 期: 卷 73, 编号 7 (2018)
- 页面: 631-640
- 栏目: Articles
- URL: https://journals.rcsi.science/1061-9348/article/view/183085
- DOI: https://doi.org/10.1134/S1061934818070055
- ID: 183085
如何引用文章
详细
New calibration equations are proposed based on the use of intensity ratios of spectral lines and peaks of coherently and incoherently scattered characteristic radiation of an X-ray tube. Prospects for their application to the X-ray fluorescence analysis of samples containing undetectable light elements are shown. The proposed method ensures high precision of determination and performance of analysis in the lack of adequate reference samples. The method was tested on an example of analysis of standard samples of bronze and samples of iron oxide materials.
作者简介
A. Garmai
Department of Chemistry
编辑信件的主要联系方式.
Email: andrew-garmay@yandex.ru
俄罗斯联邦, Moscow, 119991
K. Oskolok
Department of Chemistry
Email: andrew-garmay@yandex.ru
俄罗斯联邦, Moscow, 119991