Electrochemical reduction of selenium on a silver electrode and its determination in river water


Цитировать

Полный текст

Открытый доступ Открытый доступ
Доступ закрыт Доступ предоставлен
Доступ закрыт Только для подписчиков

Аннотация

A new, simple, and fast method is described for determining selenium based on a silver wire electrode flow-through voltammetric detector. A comparison was done between the sensitivity of disk silver, mercury-modified glassy carbon, and mercury-modified gold electrodes. The response of the flow-through voltammetric detector was evaluated with respect to each electrode’s operating potential and pH in direct current mode. The limit of detection (3s) for Se(IV) was below 0.01 mg/L. The flow-through system does not need deposition times and the organic matter does not need to be removed before determining selenium in river water.

Об авторах

Lai-Hao Wang

Department of Medical Chemistry

Автор, ответственный за переписку.
Email: e201466wang@msa.hinet.net
Тайвань, 60 Erh-Jen Road, Section 1, Jen Te, Tainan, 71743

Xi-En Shan

Department of Medical Chemistry

Email: e201466wang@msa.hinet.net
Тайвань, 60 Erh-Jen Road, Section 1, Jen Te, Tainan, 71743


© Pleiades Publishing, Ltd., 2016

Данный сайт использует cookie-файлы

Продолжая использовать наш сайт, вы даете согласие на обработку файлов cookie, которые обеспечивают правильную работу сайта.

О куки-файлах