Determination of o-phthalic acid esters in water by chromatography–mass spectrometry with emulsion dispersive liquid–liquid microextraction preconcentration


Дәйексөз келтіру

Толық мәтін

Ашық рұқсат Ашық рұқсат
Рұқсат жабық Рұқсат берілді
Рұқсат жабық Тек жазылушылар үшін

Аннотация

Using emulsion dispersive liquid–liquid microextraction preconcentration and injection of a large volume of an extract (10 μL), the limits of chromatographic–mass spectrometric detection of o-phthalic acid esters in water have been attained at the level 4 × 10–6–1 × 10–5 mg/L. The main source of the systemic error of the determination of impurities was found to be the release of o-phthalates from microparticles of chromatographic septum to the carrier gas. The extractant (n-octane) was purified by Rayleigh distillation. The independence of the concentration coefficient of the studied o-phthalates of concentration in the range (0.4–30) × 10–4 mg/L has been demonstrated. The relative expanded uncertainty of the determination of o-phthalates has been calculated and equals 12–39%.

Авторлар туралы

V. Krylov

Nizhny Novgorod State University

Хат алмасуға жауапты Автор.
Email: k658995@mail.ru
Ресей, pr. Gagarina 23, Nizhny Novgorod, 603950

V. Nesterova

Nizhny Novgorod State University

Email: k658995@mail.ru
Ресей, pr. Gagarina 23, Nizhny Novgorod, 603950


© Pleiades Publishing, Ltd., 2016