Low-Temperature Cell for Studying Absorption Spectra of Greenhouse Gases


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

A low temperature vacuum cell 220 cm in length with windows of quartz, ZnSe, and KBr has been designed for working with the high resolution Bruker IFS 125-M Fourier spectrometer. It provides a threshold sensitivity to absorption on the order of 10–7 cm–1, and allows recording absorption spectra of gases in the temperature range from 200 to 296 K in the region 1000–20 000 cm–1 with an accuracy of 0.9 K.

作者简介

V. Serdyukov

V.E. Zuev Institute of Atmospheric Optics, Siberian Branch, Russian Academy of Sciences

Email: sln@iao.ru
俄罗斯联邦, Tomsk, 634055

L. Sinitsa

V.E. Zuev Institute of Atmospheric Optics, Siberian Branch, Russian Academy of Sciences

编辑信件的主要联系方式.
Email: sln@iao.ru
俄罗斯联邦, Tomsk, 634055

A. Lugovskoi

V.E. Zuev Institute of Atmospheric Optics, Siberian Branch, Russian Academy of Sciences

Email: sln@iao.ru
俄罗斯联邦, Tomsk, 634055

N. Emelyanov

V.E. Zuev Institute of Atmospheric Optics, Siberian Branch, Russian Academy of Sciences

Email: sln@iao.ru
俄罗斯联邦, Tomsk, 634055

补充文件

附件文件
动作
1. JATS XML

版权所有 © Pleiades Publishing, Ltd., 2019