To the Precision of Measuring Concentrations of Nitroxide Radicals in Polymers by EPR Technique


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

The precision of measuring concentrations of nirtoxide radicals in polymeric matrixes using modern electron paramagnetic resonance spectrometer is analyzed. The contributions of the radical nature and concentration, signal-to-noise ratio, and the integration procedure are tested as factors influencing on the precision of the measurement. It has shown that recording spectra with modern EPR spectrometer with possibility of computer analysis of spectra, the contribution of the integration error to the total error strongly depends on the signal-to-noise ratio. If the ratio is < 20, its contribution increases fast with the decrease of this parameter. If this ratio is ≥ 20, the input of the integration error becomes negligible, and the main contribution to the total error of concentration measurements comes from errors of the Q-factor determination, which are usually close to 8–10%.

作者简介

N. Chumakova

Chemistry Department, M. V. Lomonosov Moscow State University

Email: alex-kokorin@yandex.ru
俄罗斯联邦, Moscow

T. Ivanova

Chemistry Department, M. V. Lomonosov Moscow State University

Email: alex-kokorin@yandex.ru
俄罗斯联邦, Moscow

E. Golubeva

Chemistry Department, M. V. Lomonosov Moscow State University

Email: alex-kokorin@yandex.ru
俄罗斯联邦, Moscow

A. Kokorin

N. N. Semenov Institute of Chemical Physics, Russian Academy of Sciences

编辑信件的主要联系方式.
Email: alex-kokorin@yandex.ru
俄罗斯联邦, Moscow


版权所有 © Springer-Verlag GmbH Austria, part of Springer Nature, 2018
##common.cookie##