An Optimized Passive Shimming Method for Bi-planar Permanent MRI Magnets


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

An optimized passive shimming method with iron shims is presented in this paper. First, the influence value of a single iron or magnetized shim is fast calculated and determined by analytic solution with a single practical measurement. Then, the correlation between the influence value and parameters of a single shim is analyzed, and the proper parameters, including the position, polarity, and size (radius and thickness), of the shimming pieces are well selected. Finally, the numbers and locations of the passive shims are optimized by mixed-integer linear programming method based on a modified central magnetic field. The optimized method is applied to a 0.5 T Bi-planar permanent magnet magnetic resonance imaging system, and the presented results prove the efficacy of this optimized passive shimming methodology.