Calibration of a Recorder and Remote Hydrophone in a Laboratory Tank
- Авторы: Isaev A.1, Nikolaenko A.1
-
Учреждения:
- All-Russia Research Institute of Physicotechnical and Radio Measurements (VNIIFTRI)
- Выпуск: Том 61, № 7 (2018)
- Страницы: 733-737
- Раздел: Acoustic Measurements
- URL: https://journals.rcsi.science/0543-1972/article/view/246548
- DOI: https://doi.org/10.1007/s11018-018-1493-3
- ID: 246548
Цитировать
Аннотация
We propose a method for reducing the distortion in the sensitivity calibration curve of a hydroacoustic recorder. This method is based on correction of the transmission curve for a spatial filter implemented using the complex moving weighted average (CMWA) method. This method was tested using control measurements performed using the tone-burst method, with favorable results.
Об авторах
A. Isaev
All-Russia Research Institute of Physicotechnical and Radio Measurements (VNIIFTRI)
Автор, ответственный за переписку.
Email: isaev@vniiftri.ru
Россия, Mendeleevo, Moscow Oblast
A. Nikolaenko
All-Russia Research Institute of Physicotechnical and Radio Measurements (VNIIFTRI)
Email: isaev@vniiftri.ru
Россия, Mendeleevo, Moscow Oblast