Determination of Impurities in Magnesium Alloys by Inductively Coupled Plasma Mass Spectrometry
- Authors: Alekseev A.V.1, Yakimovich P.V.1
-
Affiliations:
- State Research Center of the Russian Federation – All-Russia Research Institute of Aviation Materials (VIAM)
- Issue: Vol 62, No 8 (2019)
- Pages: 741-745
- Section: Physicochemical Measurements
- URL: https://journals.rcsi.science/0543-1972/article/view/246772
- DOI: https://doi.org/10.1007/s11018-019-01688-w
- ID: 246772
Cite item
Abstract
The methods of determination of Be, Cu, Fe, Ni, Si, Ti, and Cd in certified standard samples of magnesium alloys by inductively coupled plasma mass spectrometry (ICP-MS) are discussed. The procedure for dissolving the sample and preparing it for analysis is given. To eliminate spectral interference, a collision-reaction cell was used. An internal standard is selected and the accuracy characteristics of the magnesium alloy analysis method are calculated. The accuracy of the results is confirmed by analyzing a certified standard sample of magnesium alloy.