Enhanced Non-Flow Verification of Vortex Flow Meters


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

We analyze flow and non-flow methods for verification of vortex flow meters and describe the basic advantages and deficiencies of methods in these two broad classes. We propose an enhancement to non-flow verification of flow meters that involves simulating and recording the effect of the material flow being measured on the sensor device, and describe an experimental test bench for improved flow meter verification in which the sensor element is exposed to direct electromagnetic excitation so that it vibrates at a specified controllable frequency. The tests were performed using a Yokogawa flow meter on the experimental test bench. The results obtained confirmed that the enhanced method for verification of the vortex flow meter is feasible for practical applications.

Sobre autores

P. Startsev

Bashneft-Ufaneftekhim, Branch of Bashneft Oil Company

Autor responsável pela correspondência
Email: p.starcev@yandex.ru
Rússia, Ufa

A. Mukhamadiev

Ufa State Aviation Technical Institute

Email: p.starcev@yandex.ru
Rússia, Ufa

S. Glukhov

Bashneft-Servis Oil Refinery; UralSib Bank

Email: p.starcev@yandex.ru
Rússia, Ufa; Ufa

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Springer Science+Business Media, LLC, part of Springer Nature, 2019