Enhanced Non-Flow Verification of Vortex Flow Meters


Дәйексөз келтіру

Толық мәтін

Ашық рұқсат Ашық рұқсат
Рұқсат жабық Рұқсат берілді
Рұқсат жабық Тек жазылушылар үшін

Аннотация

We analyze flow and non-flow methods for verification of vortex flow meters and describe the basic advantages and deficiencies of methods in these two broad classes. We propose an enhancement to non-flow verification of flow meters that involves simulating and recording the effect of the material flow being measured on the sensor device, and describe an experimental test bench for improved flow meter verification in which the sensor element is exposed to direct electromagnetic excitation so that it vibrates at a specified controllable frequency. The tests were performed using a Yokogawa flow meter on the experimental test bench. The results obtained confirmed that the enhanced method for verification of the vortex flow meter is feasible for practical applications.

Авторлар туралы

P. Startsev

Bashneft-Ufaneftekhim, Branch of Bashneft Oil Company

Хат алмасуға жауапты Автор.
Email: p.starcev@yandex.ru
Ресей, Ufa

A. Mukhamadiev

Ufa State Aviation Technical Institute

Email: p.starcev@yandex.ru
Ресей, Ufa

S. Glukhov

Bashneft-Servis Oil Refinery; UralSib Bank

Email: p.starcev@yandex.ru
Ресей, Ufa; Ufa

Қосымша файлдар

Қосымша файлдар
Әрекет
1. JATS XML

© Springer Science+Business Media, LLC, part of Springer Nature, 2019