Preparation of Calibration Gas Mixtures by the Chromato-Desorption Method for Increasing the Accuracy of Quantitative Determination of Biogenic Pentane in Expired Air


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

The possibility of increasing the accuracy of preparing n-pentane calibration gas mixtures by using injection-type chromato-desorption microsystems is investigated. The possibility of obtaining calibration gas mixtures in the range of 4–12 mg/m3max = 16%) is shown. It is established that the change in the composition of the mixture during storage for three months is not more than 7%.