Preparation of Calibration Gas Mixtures by the Chromato-Desorption Method for Increasing the Accuracy of Quantitative Determination of Biogenic Pentane in Expired Air
- Authors: Platonov I.A.1, Kolesnichenko I.N.1, Novikova E.A.1, Pavlova L.V.1, Lobanova M.S.1, Mikheenkova A.E.1
-
Affiliations:
- Samara National Research University
- Issue: Vol 60, No 8 (2017)
- Pages: 848-852
- Section: Physicochemical Measurements
- URL: https://journals.rcsi.science/0543-1972/article/view/246262
- DOI: https://doi.org/10.1007/s11018-017-1281-5
- ID: 246262
Cite item
Abstract
The possibility of increasing the accuracy of preparing n-pentane calibration gas mixtures by using injection-type chromato-desorption microsystems is investigated. The possibility of obtaining calibration gas mixtures in the range of 4–12 mg/m3 (δmax = 16%) is shown. It is established that the change in the composition of the mixture during storage for three months is not more than 7%.