Mathematical Processing of Pulses for Improving the Timing Characteristics of the Correlation Counter
- Authors: Kazachek M.V.1
-
Affiliations:
- Pacific Oceanological Institute
- Issue: No 6 (2023)
- Pages: 176-180
- Section: ЛАБОРАТОРНАЯ ТЕХНИКА
- URL: https://journals.rcsi.science/0032-8162/article/view/233277
- DOI: https://doi.org/10.31857/S003281622304002X
- EDN: https://elibrary.ru/SUKBHP
- ID: 233277
Cite item
Abstract
The systematic error in measuring the duration of flashes by the correlation counter that was developed by the authors earlier on the basis of a digital oscilloscope and a computer has been reduced by 1–2 ns by mathematical processing of input pulses and by correction of their recording times. The variance in the measured duration and delay of flashes does not change when the correction is applied. The technique has been tested on model flashes, and the time of sonoluminescence flashes has been clarified. The counter can be used to measure other fast events, e.g., in nuclear physics.
About the authors
M. V, Kazachek
Pacific Oceanological Institute
Author for correspondence.
Email: tanya@poi.dvo.ru
690041, Vladivostok, Russia
References
- Казачек М.В., Гордейчук Т.В. // ПТЭ. 2019. № 1. С. 28. https://doi.org/10.1134/S0032816219010117
- Казачек М.В. Свидетельство о государственной регистрации программы для ЭВМ № 2018619059 // Опубл. 27.07. 2018.
- Walton A.J., Reynolds G.T. // Advances in Physics. 1984. V. 33. № 6. P. 595. https://doi.org/10.1080/00018738400101711
- Гордейчук Т.В., Казачек М.В. // Оптика и спектроскопия. 2020. Т. 128. № 10. С. 1492. https://doi.org/10.21883/OS.2020.10.50020.169-20
- Gompf B., Gunter R., Nick G., Pecha R., Eisenmenger W. // Physical Review Letters. 1997. V. 79. № 7. P. 1405.
- Aavikko R. Thesis …Doctor from the Helsinki University of Technology. Speciality: Science in Technology. Helsinki, Finland. 2006. P. 17.
- Ko I., Kwak H.-Y. // Journal of Physical Society of Japan. 2010. V. 79. № 12. P. 124401. https://doi.org/10.1143/JPSJ.79.124401