Time-Efficient LC/MS/MS Determination of Low Concentrations of Methylphosphonic Acid


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

A time-efficient protocol for quantification of methylphosphonic acid (MPA), a final hydrolysis product of nerve agents in aqueous environments, via liquid chromatography mass spectrometry is described. Chromatographic separation and mass spectrometry detection conditions are optimized to enable high sensitivity, selectivity, and accuracy of the approach and to eliminate impeding matrix effects associated with the analysis of water samples of natural origin. The developed technique is tested by analyzing the samples of natural waters that are spiked with a known quantity of MPA. With direct LC/MS/MS, the detection limit for MPA in natural waters is 10 ng/mL.

About the authors

T. M. Baygildiev

Moscow State University

Author for correspondence.
Email: timurbaychem@gmail.com
Russian Federation, Moscow

I. A. Rodin

Moscow State University

Email: timurbaychem@gmail.com
Russian Federation, Moscow

A. N. Stavrianidi

Moscow State University; Kostroma State Technological University

Email: timurbaychem@gmail.com
Russian Federation, Moscow; Kostroma

A. V. Braun

Moscow State University; Kostroma State Technological University

Email: timurbaychem@gmail.com
Russian Federation, Moscow; Kostroma

D. I. Akhmerova

Moscow State University

Email: timurbaychem@gmail.com
Russian Federation, Moscow

O. A. Shpigun

Moscow State University

Email: timurbaychem@gmail.com
Russian Federation, Moscow

I. V. Rybal’chenko

Kostroma State Technological University

Email: timurbaychem@gmail.com
Russian Federation, Kostroma

Supplementary files

Supplementary Files
Action
1. JATS XML