Analysis of high-purity cadmium and cadmium dioxide by atomic absorption spectrometry
- Authors: Petrova N.I.1, Lundovskaya O.V.1,2, Saprykin A.I.1,2
-
Affiliations:
- Nikolaev Institute of Inorganic Chemistry, Siberian Branch
- Novosibirsk State University
- Issue: Vol 52, No 10 (2016)
- Pages: 1020-1024
- Section: Article
- URL: https://journals.rcsi.science/0020-1685/article/view/157947
- DOI: https://doi.org/10.1134/S0020168516100113
- ID: 157947
Cite item
Abstract
To ensure high-speed analytical support to the technology of cadmium tungstate single crystal growth, we have developed an electrothermal atomic absorption spectrometry technique for analysis of high-purity cadmium and cadmium oxide for the technologically important impurities Ag, Co, Cr, Cu, Fe, Mn, Ni, Pb, and Sn with detection limits from 2 × 10–7 to 6 × 10–6 wt %.