A Device and a Method for Determining the Rate of Oxygen Consumption by an Isolated Perfused Rat Liver
- Авторлар: Shadrin K.V.1,2, Pakhomova V.G.1, Kryukova O.V.1, Rupenko A.P.1
-
Мекемелер:
- Krasnoyarsk Scientific Center, Siberian Branch, Russian Academy of Sciences
- V. F. Voino-Yasenetsky Krasnoyarsk State Medical University
- Шығарылым: Том 51, № 1 (2017)
- Беттер: 52-55
- Бөлім: Article
- URL: https://journals.rcsi.science/0006-3398/article/view/234890
- DOI: https://doi.org/10.1007/s10527-017-9683-x
- ID: 234890
Дәйексөз келтіру
Аннотация
The goal of this work was to suggest a method for measuring the total rate of oxygen consumption by an isolated perfused rat liver. This method involves calculation of the rates of oxygen consumption by the blood stream and through the liver surface. The two independently determined values are then summed up to determine the total consumption rate. Experimental results obtained by this method are in good agreement with the results obtained previously using the manometric method of measurement of the oxygen consumption rate.
Авторлар туралы
K. Shadrin
Krasnoyarsk Scientific Center, Siberian Branch, Russian Academy of Sciences; V. F. Voino-Yasenetsky Krasnoyarsk State Medical University
Хат алмасуға жауапты Автор.
Email: kvsh_buffon@mail.ru
Ресей, Krasnoyarsk; Krasnoyarsk
V. Pakhomova
Krasnoyarsk Scientific Center, Siberian Branch, Russian Academy of Sciences
Email: kvsh_buffon@mail.ru
Ресей, Krasnoyarsk
O. Kryukova
Krasnoyarsk Scientific Center, Siberian Branch, Russian Academy of Sciences
Email: kvsh_buffon@mail.ru
Ресей, Krasnoyarsk
A. Rupenko
Krasnoyarsk Scientific Center, Siberian Branch, Russian Academy of Sciences
Email: kvsh_buffon@mail.ru
Ресей, Krasnoyarsk
Қосымша файлдар