Optical biometry features in silicon oil filled eyes

封面

如何引用文章

详细

Background. The article presents results of axial length (AL) measurement in eyes filled with silicone oil and in those without silicone oil with IOLMaster and Lenstar LS 900 optical biometry methods.

Materials and methods. The anteroposterior axis was measured in 27 eyes of 27 patients with silicone oil tamponade after surgical treatment of several vitreoretinal conditions. Using IOLMaster, the AL of eyes without silicone oil tamponade varied from 21.99 mm to 29.38 mm, Lenstar LS 900 biometry gave results from 21.96 mm to 29.41 mm.

Results. According to data obtained and to their distribution, all cases were divided into 2 groups: I group — eyes with AL less than 23.63 mm, and II group — eyes with AL more than 23.63 mm. In the group II, the disparity of consecutive measurements was reliable and amounted to 0.28 ± 0.46 mm (р = 0.024) for IOLMaster and 0.23 ± 0.44 mm (р = 0.029) for Lenstar LS 900.

Conclusion. So AL values at IOLMaster and Lenstar LS 900 biometry of silicone oil filled eyes may significantly overestimate the real ones when exceeding 23.63 mm. In case of simultaneous phacoemulsification with IOL implantation, this could lead to hypermetropic shift of postoperative refraction. Lenstar LS 900 measurement error in silicon oil filled eyes is less than that of IOLMaster, thus making the first biometry method preferable. In eyes with AL shorter than 23.63 mm, the measurement difference was not reliable, thus the biometry accuracy in silicone oil filled “short” eyes becomes higher.

作者简介

Alexey Kulikov

Medical Military Academy named after S.M. Kirov

Email: pit-ark@mail.ru

MD, Professor, Head of Ophthalmology Department

俄罗斯联邦, St. Petersburg

Ekaterina Kokareva

Medical Military Academy named after S.M. Kirov

Email: pit-ark@mail.ru

MD, Head of the Hospital Department. Ophthalmology Department

俄罗斯联邦, St. Petersburg

Alexander Kuznetsov

Medical Military Academy named after S.M. Kirov

编辑信件的主要联系方式.
Email: pit-ark@mail.ru

Ophthalmology Department Resident

俄罗斯联邦, St. Petersburg, Russia

参考

  1. Захаров В.Д., Ходжаев Н.С., Горшков И.M., Mаляцинский И.A. Современная хирургия рецидива отслойки сетчатки. Обзор литературы // Офтальмология. - 2012. - Т. 9. - № 1. - С. 10-13. [Zakharov VD, Khodzhaev NS, Gorshkov IM, Malyatsinskiy IA. Current surgery of Retinal detachment recurrence. Review. Ophthalmology in Russia. 2012;9(1):10-13. (In Russ.)]. doi: 10.18008/1816-5095-2012-1-10-13.
  2. Тахчиди Х.П. Состояние эндовитреальной хирургии - реальности времени / Тезисы докладов IX Съезда офтальмологов России; Москва, 16-18 июня 2010 г. - М., 2010. - С. 232-233. [Tahchidi HP. Sostoyanie endovitreal’noy khirurgii - real’nosti vremeni. In: Proceedings of the 9th Congress of Ophthalmologists of Russia; Moskow, 16-18 Jun 2010. Moscow; 2010. P. 232-233. (In Russ.)]
  3. Riemann CD, Miller DM, Foster RE, Petersen MR. Outcomes of transconjunctival sutureless 25-gauge vitrectomy with silicone oil infusion. Retina. 2007;27(3):296-303. doi: 10.1097/01.iae.0000242761.74813.20.
  4. Shah CP, Ho AC, Regillo CD, et al. Short-term outcomes of 25-gauge vitrectomy with silicone oil for repair of complicated retinal detachment. Retina. 2008;28(5):723-728. doi: 10.1097/IAE.0b013e318166976d.
  5. Касьянов А.А., Сдобникова С.В., Троицкая Н.А., Рыжкова Е.Г. Расчёт оптической силы интраокулярной линзы у пациентов с силиконовой тампонадой // Вестник офтальмологии. - 2015. - Т. 131. - № 5. - С. 26-31. [Kas’yanov AA, Sdobnikova SV, Troitskaya NA, Ryzhkova EG. Intraocular lens power calculation in silicone-filled eyes. Annals of ophtalmology. 2015;131(5):26-31. (In Russ.)]. doi: 10.17116/oftalma2015131526-31.
  6. Столяренко Г.E., Сдобникова С.В. Современное состояние трансвитреальной хирургии глаза // Вестник Российской академии медицинских наук. - 2003. - № 2. - С. 15-20. [Stolyarenko GE, Sdobnikova SV State-of-art of endo-and transvitreal surgery of the eye. Annals of the Russian Academy of Medical Sciences. 2003;(2):15-20. (In Russ.)]
  7. Юодкайте Г.Ю. Изменение тканей глаза при введении силикона в стекловидное тело // Офтальмологический журнал. - 1971. - Т. 26. - № 2. - С. 96-98. [Yuodkayte GY. Izmenenie tkaney glaza pri vvedenii silikona v steklovidnoe telo. Oftalmol Zh. 1971;26(2):96-98. (In Russ.)]
  8. Batra A, Vemuganti GK, Das T, et al. Does Silicone Oil Penetrate the Lens Capsule? Retina. 2001;21(3):275-277. doi: 10.1097/00006982-200106000-00019.
  9. Oner HE, Durak I, Saatci OA. Phacoemulsification and foldable intraocular lens implantation in eyes filled with silicone oil. Ophthalmic Surg Lasers Imaging. 2003;34(5):358-362. doi: 10.3928/1542-8877-20030901-03.
  10. Tanner V, Haider A, Rosen P. Phacoemulsification and combined management of intraocular silicone oil. J Cataract Refract Surg. 1998;24(5):585-591. doi: 10.1016/s0886-3350(98)80250-2.
  11. Аванесова Т.А. Регматогенная отслойка сетчатки: современное состояние проблемы // Офтальмология. - 2015. - Т. 12. - № 1. - С. 24-32. [Avanesova TA. Rhegmatogenous retinal detachment: current opinion. Ophthalmology in Russia. 2015;12(1):24-32. (In Russ.)]. doi: 10.18008/1816-5095-2015-1-24-32.
  12. Gharbiya M, Grandinetti F, Scavella V, et al. Correlation between Spectral-Domain Optical Coherence Tomography Findings and Visual Outcome after Primary Rhegmatogenous Retinal Detachment Repair. Retina. 2012;32(1):43-53. doi: 10.1097/IAE.0b013e3182180114.
  13. Wakabayashi T, Oshima Y, Fujimoto H, et al. Foveal microstructure and visual acuity after retinal detachment repair: imaging analysis by Fourier-domain optical coherence tomography. Ophthalmology. 2009;116(3):519-528. doi: 10.1016/j.ophtha.2008.10.001.
  14. Gullstrand A. Die Optische Abbildung in heterogen Medien die Dioptrik der Kristallince des Menschen. K Sven Vetenskapsakad Handl. 1908;43:1-32. (In German).
  15. Даниленко Е.В. Оптимизация расчёта оптической силы интраокулярной линзы, имплантируемой при факоэмульсификации: Дис. … канд. мед. наук. - СПб., 2012. [Danilenko EV. Optimizatsiya rascheta opticheskoy sily intraokulyarnoy linzy, implantiruemoy pri fakoemul’sifikatsii. [dissertation] Saint Petersburg; 2012. (In. Russ.)]
  16. Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238(9):765-773. doi: 10.1007/s004170000188.
  17. Dietlein TS, Roessler G, Luke C, et al. Signal quality of biometry in silicone oil-filled eyes using partial coherence laser interferometry. J Cataract Refract Surg. 2005;31(5):1006-1010. doi: 10.1016/j.jcrs.2004.09.049.