Medical phantom of the knee joint for computed tomography studies

Cover Page

Cite item

Full Text

Abstract

BACKGROUND: The knee joint is a frequently visualized anatomical region in clinical practice. Accurate interpretation of CT scans necessitates a comprehensive understanding of anatomy and a sound grasp of fundamental technical principles and imaging protocols. To safeguard the patient's well-being, it is of paramount importance to prevent erroneous studies resulting from suboptimal equipment quality, setup issues, and patient positioning. These difficulties can be circumvented by the use of phantoms to pre-adjust the equipment and the provision of training to medical staff in scanning techniques.

AIM: The aim of the study was to develop a technique for creating an anthropomorphic medical phantom of the knee joint that would accurately reflect the X-ray density of the corresponding human tissues, thus enabling the use of computed tomography studies.

MATERIALS AND METHODS: The knee joint phantom comprises a series of models representing the femur, tibia, fibula, patella, collateral ligaments, lateral and medial menisci, tendon of the quadriceps femoris muscle, anterior and posterior cruciate ligaments, and patellar ligament. Ligament models were 3D-printed from resin, bones were cast from silicone, soft tissues were modeled with a homogeneous structure of silicone-like materials and made by casting into silicone molds. The skin was similarly modeled. In the study, the anode voltage range of the CT scanner varied from 80 to 140 kV, and the slice thickness was equal to 1.25 mm.

RESULTS: The developed anthropomorphic knee joint phantom demonstrated the X-ray density of the modeled anatomical structures, with ligaments exhibiting a range of 80–120 units on the Hausfield scale, bones exhibiting a range of 320–370 units, and soft tissues and skin exhibiting a range of 20–60 units. The use of additive technologies made it possible to achieve a high degree of similarity between the phantom forms and the knee joint. Further research may be directed towards the creation of a more complex model of bone tissue, comprising a separate cortical layer and spongy substance.

CONCLUSIONS: The use of an anthropomorphic knee phantom allows for the acquisition of high-quality CT images without the need for prior scanning of patients.

About the authors

Ekaterina D. Belyakova

Research and Practical Clinical Center for Diagnostics and Telemedicine Technologies

Email: belyakova_e_d@student.sechenov.ru
ORCID iD: 0009-0009-0316-0628
Russian Federation, Moscow

Anastasia A. Nasibullina

Research and Practical Clinical Center for Diagnostics and Telemedicine Technologies

Email: NasibullinaAA@zdrav.mos.ru
ORCID iD: 0000-0003-1695-7731
SPIN-code: 2482-3372
Russian Federation, Moscow

Julia V. Bulgakova

Research and Practical Clinical Center for Diagnostics and Telemedicine Technologies

Email: BulgakovaYV@zdrav.mos.ru
ORCID iD: 0000-0002-1627-6568
SPIN-code: 8945-6205
Russian Federation, Moscow

Olga V. Vlasova

Research and Practical Clinical Center for Diagnostics and Telemedicine Technologies

Author for correspondence.
Email: VlasovaOV10@zdrav.mos.ru
ORCID iD: 0000-0002-4211-9566
SPIN-code: 4492-3864
Russian Federation, Moscow

Veronika V. Grebennikova

Research and Practical Clinical Center for Diagnostics and Telemedicine Technologies

Email: GrebennikovaVV@zdrav.mos.ru
ORCID iD: 0009-0003-6041-132X
SPIN-code: 4949-1057
Russian Federation, Moscow

Olga V. Omelyanskaya

Research and Practical Clinical Center for Diagnostics and Telemedicine Technologies

Email: OmelyanskayaOV@zdrav.mos.ru
ORCID iD: 0000-0002-0245-4431
SPIN-code: 8948-6152
Russian Federation, Moscow

Alexey V. Petraikin

Research and Practical Clinical Center for Diagnostics and Telemedicine Technologies

Email: alexeypetraikin@gmail.com
ORCID iD: 0000-0003-1694-4682
SPIN-code: 6193-1656

MD, Dr. Sci. (Med.)

Russian Federation, Moscow

Denis V. Leonov

Research and Practical Clinical Center for Diagnostics and Telemedicine Technologies

Email: LeonovDV2@zdrav.mos.ru
ORCID iD: 0000-0003-0916-6552
SPIN-code: 5510-4075

Cand. Sci. (Tech.)

Russian Federation, Moscow

References

  1. Kapisiz A, Kaya C, Eryilmaz S, et al. Letter to the Editor in Response to: Reducing Unnecessary Computed Tomography Scan. Journal of pediatric surgery. 2023;58(7):1402. doi: 10.1016/j.jpedsurg.2023.02.048
  2. Hernandez AM, Bayne CO, Bateni C, et al. Extremity radiographs derived from low-dose ultra-high-resolution CT: a phantom study. Skeletal Radiol. 2024. doi: 10.1007/s00256-024-04600-y
  3. Leonov D, Venidiktova D, Costa-Júnior JFS, et al. Development of an anatomical breast phantom from polyvinyl chloride plastisol with lesions of various shape, elasticity and echogenicity for teaching ultrasound examination. Int J CARS. 2023;19:151–161. doi: 10.1007/s11548-023-02911-4
  4. Leonov D, Kodenko M, Leichenco D, Nasibullina A, Kulberg N. Design and validation of a phantom for transcranial ultrasonography. Int J Comput Assist Radiol Surg. 2022;17:1579–1588. doi: 10.1007/s11548-022-02614-2

Copyright (c) 2024 Eco-Vector

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

This website uses cookies

You consent to our cookies if you continue to use our website.

About Cookies