Uncooled Microbolometer Based on VOx Film for Infrared and Millimeter-Wave Radiation


Дәйексөз келтіру

Толық мәтін

Ашық рұқсат Ашық рұқсат
Рұқсат жабық Рұқсат берілді
Рұқсат жабық Тек жазылушылар үшін

Аннотация

The construction of an uncooled infrared and millimeter-wave microbolometer has been proposed. The microbolometer consists of a converter structure inside a metal box with an input glass window. The converter structure consists of a thin mica substrate covered with a frequency-dependent metal surface on the front side and thermosensitive VOx elements with an electric circuit on the back side. The absorption of infrared radiation and millimeter waves is investigated. Radiation detection is studied numerically and experimentally in ranges of 0.4–16.66 μm and (0.33–0.37) × 103 μm. The exploitation parameters of the device are determined and areas of application for the device are proposed.

Авторлар туралы

A. Oleynik

Yuri Gagarin State Technical University of Saratov

Email: medmixal@yahoo.com
Ресей, Saratov, 410054

M. Medvedev

Yuri Gagarin State Technical University of Saratov

Хат алмасуға жауапты Автор.
Email: medmixal@yahoo.com
Ресей, Saratov, 410054

Y. Turkin

Yuri Gagarin State Technical University of Saratov

Email: medmixal@yahoo.com
Ресей, Saratov, 410054

Қосымша файлдар

Қосымша файлдар
Әрекет
1. JATS XML

© Pleiades Publishing, Ltd., 2017