Procedure for Measuring the Mass Fractions of Products of the Catalytic Ethylbenzene Pyrolysis Reaction


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

A technique is proposed for determining the mass fractions of products of the catalytic ethylbenzene pyrolysis reaction on a capillary column with functionalized poly(1-trimethylsilyl-1-propyne) (PTMSP/N2O). The capillary column with PTMSP/N2O selectively separates methane, ethane, ethylene, and acetylene (light C1–C2 hydrocarbons), along with benzene, toluene, ethylbenzene, and styrene (aromatic hydrocarbons). A procedure is developed to measure the mass fractions of light (C1–C2) and aromatic hydrocarbons in the gas phase. The ranges of analytical measurement are 2.9 × 10−8 to 1.2 × 10−1 mg/mL for light C1–C2 hydrocarbons and 3.5 × 10−11 to 4.0 × 10−3 mg/mL for liquid hydrocarbons. The under conditions of repeatability is 1.9–4.7%.