Practical Aspects of Mastering the Verification of Temperature Standards of the 1st Category
- Authors: Golobokov M.V.1
-
Affiliations:
- State Regional Center for Standardization, Metrology and Testing in the Novosibirsk Region
- Issue: No 2 (2025)
- Pages: 42-47
- Section: Metrology
- URL: https://journals.rcsi.science/1993-8780/article/view/296177
- DOI: https://doi.org/10.24412/1993-8780-2025-2-42-47
- ID: 296177
Cite item
Abstract
I have described the solution of a number of problems that arose during the development of verification of temperature standards of the 1st category. A refined procedure for determining inhomogeneity is proposed, additional requirements for external conditions during measurements are listed, and the permissible value of inhomogeneity is specified. It has been experimentally confirmed that the method established in the verification procedure for estimating the random component of the measurement error when verifying reference thermoelectric thermal converters of the 1st category underestimates the results obtained by 1.2–2.1 times.
The work can be useful for specialists involved in the verification of measuring instruments and temperature standards.
Full Text

About the authors
M. V. Golobokov
State Regional Center for Standardization, Metrology and Testing in the Novosibirsk Region
Author for correspondence.
Email: malachit_40@inbox.ru
PhD (Tech.), Leading Metrology Engineer of the Thermal Engineering Measurements Department
Russian Federation, NovosibirskReferences
- MKSN.681118.013 RE Furnace of reference points of PRT-600-1100-2: User manual, Omsk, NPP Etalon, 2018.
- Freeze Point Furnace: User manual, USA, Fluke corp., 2018.
- GOST R 8.611–2024 SSM. Thermoelectric converters platinumrhodium-platinum reference of the 1st, 2nd and 3rd grades. Verification and calibration procedure.
- Uncertainty of calibration of a thermocouple by comparison with a sample thermocouple in a furnace; https://temperatures.ru/pages/neopredelennost_kalibrovki_termopary (acc.: 27.09.2024).
- GOST 8.461–2009 SSM. Thermal resistance converters made of platinum, copper and nickel. Verification procedure.
- GOST R 8.585–2001 SSM. Thermocouples. Nominal static conversion characteristics.
- Guidelines on the Calibration of Thermocouples EURAMET Calibration Guide N 8 Version 3.1, 2020; https://www.euramet.org/publications-media-centre/calibration-guidelines (acc.: 27.09.2024).
- Golobokov M.V. Software Rubin, Library Standard. Certificate of state registration of the computer program оf 6.11.2019 N 2019665207.
- GOST R 8.571–98 SSM. Resistance thermometers platinum reference of the 1st and 2nd grades. Verification procedure.
Supplementary files