Experimental study of the effects of laser sialolithotripsy
- 作者: Zhuchkova D.1,2, Sysolyatin S.1,2,3
-
隶属关系:
- Peoples’ Friendship University of Russia
- “Endostom” clinic
- Medical and Biological University of Innovation and Continuing Education of the Burnazyan Federal Medical Biophysical Center
- 期: 卷 27, 编号 5 (2023)
- 页面: 395-401
- 栏目: Experimental and Theoretical Investigation
- URL: https://journals.rcsi.science/1728-2802/article/view/251368
- DOI: https://doi.org/10.17816/dent346678
- ID: 251368
如何引用文章
详细
BACKGROUND: The effectiveness of removing concretions using sialoendoscopy depends on the size and location of the concretion in the salivary gland. According to the research results, holmium YAG lasers are the predominant intracorporeal lithotriptors in crushing urolites and sialoliths. However, to date, in comparative studies of lasers of different types in urology, the Thule laser FiberLase U2 demonstrates high efficiency. Research by urologist surgeons served as the basis for the experiment on the possibility of crushing sialoliths and analyzing the physical effects.
AIM: To evaluate the safety and time of crushing of sialolith using a thulium laser FiberLase U2 in vitro.
MATERIALS AND METHODS: Twelve round-shaped sialolith with a diameter of 5–6 mm were selected for the study. A model simulating the gland duct was developed. The fiber of the FiberLase U2 laser was fed through the proximal end of the model, and a series of lithotripsies were performed sequentially in three modes of the device at maximum and minimum values of energy and pulse frequency, fixing the temperature rise during irrigation and the resulting effects during crushing.
RESULTS: Crushing sialoliths to fragments measuring <1 mm is possible in all three laser modes. Operation at maximum values of pulse energy and frequency proceeded faster, whereas the maximum temperature rise varied from 45°C to 48°C, depending on the mode. Crushing was accompanied by pronounced cavitation and frequent sparks when the laser pulse was applied. Fragmentation at the minimum parameters of the pulse energy and frequency proceeded much slower; however, the maximum temperature rise ranged from 33°C to 40°C, and the listed effects were observed to a lesser extent in all three modes.
CONCLUSION: The use of a Thule laser for crushing concretions in all three modes at maximum pulse energy and frequency is inefficient and unsafe. Work at minimum values of the dusting and popcorning modes has shown the best results; however, further study of crushing with this laser is necessary.
作者简介
Darya Zhuchkova
Peoples’ Friendship University of Russia; “Endostom” clinic
编辑信件的主要联系方式.
Email: DaLitvinova@mail.ru
ORCID iD: 0000-0002-9200-4257
SPIN 代码: 4679-9403
Assistant
俄罗斯联邦, 6 Miklukho-Maklaya street, 117198 Moscow; MoscowSvyatoslav Sysolyatin
Peoples’ Friendship University of Russia;“Endostom” clinic; Medical and Biological University of Innovation and Continuing Education of the Burnazyan Federal Medical Biophysical Center
Email: sp-sysolyatin@yandex.ru
ORCID iD: 0000-0002-5794-9087
SPIN 代码: 2050-5215
MD, Dr. Sci. (Med.), Professor
俄罗斯联邦, 6 Miklukho-Maklaya street, 117198 Moscow; Moscow; Moscow参考
- Sysolyatin SP, Bannikova KA, Sysolyatin PG, Gaytova VG, Baydik OD. Endosialoscopic diagnosis and treatment of sialolithiasis. The Siberian Scientific Medical Journal. 2020;40(1):45–52. doi: 10.15372/SSMJ20200106
- Oddon PA, Royer G, Graillon N, et al. Treatment of salivary stones by intraductal pneumatic lithotripsy: a preliminary presentation of the StoneBreaker with sterile bag covering. J Stomatol Oral Maxillofac Surg. 2017;118(2):119–121. doi: 10.1016/j.jormas.2017.02.001
- Strychowsky JE, Sommer DD, Gupta MK, Cohen N, Nahlieli O. Sialendoscopy for the management of obstructive salivary gland disease: a systematic review and meta-analysis. Arch Otolaryngol Head Neck Surg. 2012;138(6):541–547. doi: 10.1001/archoto.2012.856
- Galdermans M, Gemels B. Success rate and complications of sialendoscopy and sialolithotripsy in patients with parotid sialolithiasis: a systematic review. Oral Maxillofac Surg. 2020;24(2):145–150. doi: 10.1007/s10006-020-00834-x
- Koch M, Hung SH, Su CH, et al. Intraductal lithotripsy in sialolithiasis with two different Ho:YAG lasers: presetting parameters, effectiveness, success rates. Eur Rev Med Pharmacol Sci. 2019;23(13):5548–5557. doi: 10.26355/eurrev_201907_18288
- Faklaris I, Bouropoulos N, Vainos NA. Sialolithiasis: Application parameters for an optimal laser therapy. J Biophotonics. 2020;13(7):e202000044. doi: 10.1002/jbio.202000044
- Koch M, Schapher M, Mantsopoulos K, Iro H. Intraductal lithotripsy in sialolithiasis using the Calculase III™ Ho:YAG laser: first experiences. Lasers Surg Med. 2021;53(4):488–498. doi: 10.1002/lsm.23325
- Schrötzlmair F, Müller M, Pongratz T, et al. Laser lithotripsy of salivary stones: Correlation with physical and radiological parameters. Lasers Surg Med. 2015;47(4):342–349. doi: 10.1002/lsm.22333
- Martellucci S, Pagliuca G, de Vincentiis M, et al. Ho:Yag laser for sialolithiasis of Wharton’s duct. Otolaryngol Head Neck Surg. 2013;148(5):770–774. doi: 10.1177/0194599813479914
- Capaccio P, Torretta S, Pignataro L, Koch M. Salivary lithotripsy in the era of sialendoscopy. Acta Otorhinolaryngol Ital. 2017;37(2):113–121. doi: 10.14639/0392-100X-1600
- Traxer O, Keller EX. Thulium fiber laser: the new player for kidney stone treatment? A comparison with Holmium:YAG laser. World J Urol. 2020;38(8):1883–1894. doi: 10.1007/s00345-019-02654-5
- Kałużny J, Klimza H, Tokarski M, et al. The holmium:YAG laser lithotripsy—a non-invasive tool for removal of midsize stones of major salivary glands. Lasers Med Sci. 2020;37(1):163–169. doi: 10.1007/s10103-020-03201-0
- Martov AG, Baranov AV, Biktimirov RG, Alpin DM, Biktimirov TR. Application of laser radiation in urology. Lazernaya medicina. 2020;24(1):57–62. doi: 10.37895/2071-8004-2020-24-1-57-62
- Martov AG, Ergakov DV, Guseinov MA, et al. Initial experience in clinical application of thulium laser contact lithotripsy for transurethral treatment of urolithiasis. Urologiia. 2018;(1):112–120. doi: 10.18565/urology.2018.1.112-120
- Keller EX, De Coninck V, Doizi S, Daudon M, Traxer O. Thulium fiber laser: ready to dust all urinary stone composition types. World J Urol. 2021;39(6):1693–1698. doi: 10.1007/s00345-020-03217-9
- Blackmon RL, Irby PB, Fried NM. Comparison of holmium:YAG and thulium fiber laser lithotripsy: ablation thresholds, ablation rates, and retropulsion effects. J Biomed Opt. 2011;16(7):071403. doi: 10.1117/1.3564884
- Hardy LA, Wilson CR, Irby PB, Fried NM. Thulium fiber laser lithotripsy in an in vitro ureter model. J Biomed Opt. 2014;19(12):128001. doi: 10.1117/1.JBO.19.12.128001
- Popov SV, Orlov IN, Sytnik DA, et al. Thulium and holmium ureterolithotripsy: evaluation of thermal effects on the ureter by measuring the temperature of the irrigation fluid in vitro. Experimental and Clinical Urology. 2021;14(1):26–30. doi: 10.29188/2222-8543-2021-14-1-26-30
- Sevostyanova OA, Boshchenko VS, Osadchii VK, Parnachev VP, Polienko AK. The study of mineral composition and structure of uroliths in the residents of Tomsk district (Tomsk). Urologiia. 2017;(2):76–81. doi: 10.18565/urol.2017.2.76-81
- Kraaij S, Brand HS, van der Meij EH, de Visscher JG. Biochemical composition of salivary stones in relation to stone-and patient-related factors. Med Oral Patol Oral Cir Bucal. 2018;23(5):e540–e544. doi: 10.4317/medoral.22533
- Durbec M, Dinkel E, Vigier S, et al. Thulium-YAG laser sialendoscopy for parotid and submandibular sialolithiasis. Lasers Surg Med. 2012;44(10):783–786. doi: 10.1002/lsm.22094
- Zhuchkova DV, Sysolyatin SP. Experimental investigation of the effect of retropulsion in sialolithotripsy with a thulium laser. Clinical Dentistry. 2023;26(1):121–125. doi: 10.37988/1811-153X_2023_1_121