Nanosecond Calibration of a Piezo Transducer by Comparing Thermoacoustic and Vaporization Pressure Signals at Pulsed Laser Irradiation of a Metal Target
- Авторлар: Samokhin A.A.1, Shashkov E.V.1, Vorobiev N.S.1, Zubko A.E.1
-
Мекемелер:
- Prokhorov General Physics Institute of the Russian Academy of Sciences
- Шығарылым: Том 27, № 4 (2019)
- Беттер: 268-270
- Бөлім: Interaction of Laser Radiation with Matter
- URL: https://journals.rcsi.science/1541-308X/article/view/218278
- DOI: https://doi.org/10.3103/S1541308X19040046
- ID: 218278
Дәйексөз келтіру
Аннотация
A possibility of calibrating a piezo transducer by comparing thermoacoustic and vaporization pressure signals generated in a metal (liquid mercury) irradiated with nanosecond laser pulses is demonstrated. In this calibration method, significant difference in the dependence of these two pressure generation mechanisms on the irradiated surface temperature is used.
Авторлар туралы
A. Samokhin
Prokhorov General Physics Institute of the Russian Academy of Sciences
Хат алмасуға жауапты Автор.
Email: asam40@mail.ru
Ресей, ul. Vavilova 38, Moscow, 119991
E. Shashkov
Prokhorov General Physics Institute of the Russian Academy of Sciences
Хат алмасуға жауапты Автор.
Email: shashkov@kapella.gpi.ru
Ресей, ul. Vavilova 38, Moscow, 119991
N. Vorobiev
Prokhorov General Physics Institute of the Russian Academy of Sciences
Хат алмасуға жауапты Автор.
Email: vor@kapella.gpi.ru
Ресей, ul. Vavilova 38, Moscow, 119991
A. Zubko
Prokhorov General Physics Institute of the Russian Academy of Sciences
Хат алмасуға жауапты Автор.
Email: aleksey.zubko11@gmail.com
Ресей, ul. Vavilova 38, Moscow, 119991
Қосымша файлдар