Medical gamma probe based on LaBr3:Ce scintillation crystal


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

The medical gamma probe for radionuclide diagnostics based on LaBr3:Ce scintillation crystal and a silicon photomultiplier is described. Spectrometric characteristics of the detector are studied. The energy resolution is measured to be 3.6%at 662 keV and 10% at 140.5 keV. Technical parameters of the developed gamma probe are described. It is shown that it matches the best world analogues in the set of characteristics.

Sobre autores

A. Berdnikova

National Research Nuclear University “MEPhI”; Lebedev Physical Institute

Autor responsável pela correspondência
Email: yagn.anastaia@gmail.com
Rússia, Kashirskoe sh. 31, Moscow, 115409; Leninskii pr. 53, Moscow, 119991

F. Dubinin

Lebedev Physical Institute

Email: yagn.anastaia@gmail.com
Rússia, Leninskii pr. 53, Moscow, 119991

V. Kantserov

Lebedev Physical Institute

Email: yagn.anastaia@gmail.com
Rússia, Leninskii pr. 53, Moscow, 119991

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Allerton Press, Inc., 2016