Medical gamma probe based on LaBr3:Ce scintillation crystal
- Autores: Berdnikova A.K.1,2, Dubinin F.A.2, Kantserov V.A.2
-
Afiliações:
- National Research Nuclear University “MEPhI”
- Lebedev Physical Institute
- Edição: Volume 43, Nº 8 (2016)
- Páginas: 249-251
- Seção: Article
- URL: https://journals.rcsi.science/1068-3356/article/view/228031
- DOI: https://doi.org/10.3103/S1068335616080042
- ID: 228031
Citar
Resumo
The medical gamma probe for radionuclide diagnostics based on LaBr3:Ce scintillation crystal and a silicon photomultiplier is described. Spectrometric characteristics of the detector are studied. The energy resolution is measured to be 3.6%at 662 keV and 10% at 140.5 keV. Technical parameters of the developed gamma probe are described. It is shown that it matches the best world analogues in the set of characteristics.