Spectrometer–telescope calibration by cosmic muons
- Authors: Baskov V.A.1, Govorkov B.B.1, Polyanskii V.V.1
-
Affiliations:
- Lebedev Physical Institute
- Issue: Vol 43, No 3 (2016)
- Pages: 108-111
- Section: Article
- URL: https://journals.rcsi.science/1068-3356/article/view/227898
- DOI: https://doi.org/10.3103/S1068335616030052
- ID: 227898
Cite item
Abstract
The results of cosmic-muon calibration of the multichannel lead–scintillation “sandwich”-type spectrometer–telescope intended for operation in high-intensity photon and electron beams, capable of determining the beam energy and forming intrinsic trigger, are presented. It was found that the relative energy resolution at an average energy release by cosmic muons in the spectrometer of ∼26 MeV is from 7.5% to 23% depending on the internal and external trigger version.
About the authors
V. A. Baskov
Lebedev Physical Institute
Author for correspondence.
Email: baskov@x4u.mail.lebedev.ru
Russian Federation, Leninskii pr. 53, Moscow, 119991
B. B. Govorkov
Lebedev Physical Institute
Email: baskov@x4u.mail.lebedev.ru
Russian Federation, Leninskii pr. 53, Moscow, 119991
V. V. Polyanskii
Lebedev Physical Institute
Email: baskov@x4u.mail.lebedev.ru
Russian Federation, Leninskii pr. 53, Moscow, 119991
Supplementary files