Intricacies of the Determination of the Radiochemical Purity of 68Ga Preparations: Possibility of Sorption of Ionic 68Ga Species on Reversed-Phase Columns


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

The results of studying 68Ga radiopharmaceuticals using various TLC and HPLC procedures are compared. The data obtained reliably show that a part of 68Ga ionic species are irreversibly (under definite conditions) sorbed onto chromatographic columns packed with С18 reversed phase. The loss of 68Ga ionic species in the analysis can reach 90%. The 68Ga loss increases with an increase in pH of the preparation. At pH 2.5–3.0, the total loss of 68Ga ionic species on the chromatographic column does not exceed 15%. At pH 4.0, it is 65 ± 7% on the average, and at pH 6.0 it reaches 87 ± 8%. This effect should be taken into account in analysis of any 68Ga radiopharmaceuticals.

作者简介

A. Larenkov

State Research Center Burnasyan Federal Medical Biophysical Center

编辑信件的主要联系方式.
Email: anton.larenkov@gmail.com
俄罗斯联邦, Zhivopisnaya ul. 46, Moscow, 123182

A. Maruk

State Research Center Burnasyan Federal Medical Biophysical Center

Email: anton.larenkov@gmail.com
俄罗斯联邦, Zhivopisnaya ul. 46, Moscow, 123182

G. Kodina

State Research Center Burnasyan Federal Medical Biophysical Center

Email: anton.larenkov@gmail.com
俄罗斯联邦, Zhivopisnaya ul. 46, Moscow, 123182


版权所有 © Pleiades Publishing, Inc., 2018