Calculation of the Dose Distribution Near the Gantry of a Radiotherapy System
- Autores: Bondarenko T.V.1, Polikhov S.A.1, Smirnov V.P.1, Kurilik A.S.2, Ovchinnikova L.Y.2
-
Afiliações:
- Research Institute of Technical Physics and Automation (NIITFA)
- Research Institute of Nuclear Physics (NIIYaF MGU) and Laboratory of Electron Accelerators (LEU MGU), Lomonosov Moscow State University
- Edição: Volume 125, Nº 5 (2019)
- Páginas: 338-344
- Seção: Article
- URL: https://journals.rcsi.science/1063-4258/article/view/249141
- DOI: https://doi.org/10.1007/s10512-019-00491-8
- ID: 249141
Citar
Resumo
Data on the radiation loads of a radiotherapy system developed at NIITFA are presented. The main problems of the calculations are to establish the region in the mobile part of the system that is best suited for arrangement of blocks with electronics and to calculate the parameters of local radiation shielding that ensures fault-free operation of the electronics for the life of the system. The calculations were performed by the Monte Carlo method using the FLUKA software package.
Sobre autores
T. Bondarenko
Research Institute of Technical Physics and Automation (NIITFA)
Autor responsável pela correspondência
Email: maryjane.gargar@springer.com
Rússia, Moscow
S. Polikhov
Research Institute of Technical Physics and Automation (NIITFA)
Email: maryjane.gargar@springer.com
Rússia, Moscow
V. Smirnov
Research Institute of Technical Physics and Automation (NIITFA)
Email: maryjane.gargar@springer.com
Rússia, Moscow
A. Kurilik
Research Institute of Nuclear Physics (NIIYaF MGU) and Laboratory of Electron Accelerators (LEU MGU), Lomonosov Moscow State University
Email: maryjane.gargar@springer.com
Rússia, Moscow
L. Ovchinnikova
Research Institute of Nuclear Physics (NIIYaF MGU) and Laboratory of Electron Accelerators (LEU MGU), Lomonosov Moscow State University
Email: maryjane.gargar@springer.com
Rússia, Moscow
Arquivos suplementares