Ways of Developing Analyzers for Static Mass Spectrometers
- Authors: Spivak-Lavrov I.F.1, Baisanov O.A.2, Nurmukhanova A.A.1
-
Affiliations:
- Zhubanov Aktobe Regional State University
- Begel’dinov Military Institute, Air Defense Force
- Issue: Vol 82, No 10 (2018)
- Pages: 1353-1358
- Section: Article
- URL: https://journals.rcsi.science/1062-8738/article/view/186597
- DOI: https://doi.org/10.3103/S1062873818100210
- ID: 186597
Cite item
Abstract
A prism mass spectrometer with triple focusing of the ion beam using a cone-shaped achromatic prism (CSAP) is considered. The CSAP has a record high angular mass dispersion of approximately 50 rad upon 100% variation of the mass. The performance of a prism mass analyzer with collimating and focusing transaxial lenses is simulated using the Monte Carlo method. The resolution and the quality parameters of such a mass analyzer are shown to exceed those of existing static devices by several times. A resolving power of several million can be obtained using narrow beams on a small device with sub-meter dimensions.